laboratory equipment

Our modern measurement equipment includes high-quality devices for e. g.:

 

The full-absorption hall

The heart of the EMC laboratory is the large full-absorber hall with hybrid absorber lining. It can be used to perform a wide range of EMC measurements on your DUT.

Feasible tests in the full absorber hall

  • Standardized immunity tests according to IEC (EN) 61000-4-3
  • Standard-compliant emission measurements according to CISPR 16-1-4
  • Standardized test and measurement procedures for electromagnetic compatibility according to EN 61000-4-22
  • Emission measurements 30 MHz – 6 GHz
  • Immunity tests 80 MHz – 1 GHz with 20 V/m and
    1 GHz – 6 GHz with 10 V/m, higher levels can also be set during development

Equipment and dimensions

  • Automatically controlled test piece turntable
  • Remote-controlled camera system for optical and acoustic observation of the test specimen
  • Automatic polarization switching of the antennas
  • Testing of electropneumatic DUTs
  • Copper granulate chute for decoupling shielded cables
  • Filtered analog signals and galvanically isolated frequency
  • EMC controller, signal generator, amplifier, switching matrix and directional coupler
  • Max. Dimensions of DUT: 1,500 mm x 2,000 mm (W x H)
    – Weight up to 500 kg
  • Measuring receiver Rhode & Schwarz ESR 7, frequency range:
    9 KHz – 7 GHz

Antennas

For emission measurements

AXL-4000E Logarithmic per. Wideband antenna
Frequency range: 25 MHz – 4 GHz

For noise immunity measurements

AXL-80 Stacked log. per. Wideband antenna
Frequency range: 80 MHz – 4 GHz

For emission and immunity measurements

MAX9 Stacked log. per. Wideband antenna
Frequency range: 0.7 GHz – 10.5 GHz

Modern measuring equipment

Emission measurements- Conducted

Current clamp

9 KHz – 30 MHz
Capacitive coupling clamp for signal lines

TK 9420 probe

9 KHz – 30 MHz
Application with high current consumption

Mesh replica

  • Three-phase, 32 A complies with CISPR 16
  • Frequency range: 9 KHz – 30 MHz
  • Phases: 3+N
  • man. and autom. phase switching
  • Max. AC voltage 400 V and 565 V DC voltage
  • Current: 4 x 32 A
  • Several 2-pole mains simulations for 9 KHz – 30 MHz

Emission measurements - flicker and harmonics measurements

DPA 500N

  • Real-time data recording
  • Internal hard disk for data storage
  • 16-bit A/D converter
  • Wide range current input
  • Built-in flicker impedance

Netwave 7-400

  • High power bandwidth: DC – 5 kHz
  • Output power up to 7.5 kVA AC,
    9.0 kW DC
  • Output voltage 360 VAC, or +/- 500 VDC
  • High peak current capability up to 200 A

Conducted high-frequency injection

TESEQ CDN

  • CDN M532 – 1000 V
  • CDN T2A-10
  • CDN T4A-10
  • KEMZ801A

AMETEK NSG4070

Multifunctional device for accompanying development and conformity tests

Immunity measurements - conducted

ESD Burst-Surge Generator

  • For simulation of static discharges, indirect lightning effects, magnetic field measurements, voltage reduction, pulse heights and bouncing of mechanical contacts
  • Signal coupling filter CDN-UTP ED3
  • Three-phase coupling filter CD 2000-6-32

Contact:

Josip Horvat
Division Manager EMC Laboratory
Fon: +49 (7172) 926 13-33
E-Mail: josip.horvat@baudisch.de

Carsten Hampel
EMC-Expert
Fon: +49 (7172) 926 13-34
E-Mail: carsten.hampel@baudisch.de

 